CVEvulnerability

CVE-2014-5432

9.8 CRITICAL Published 2019-03-26 Modified 2026-06-17 Official page ↗
Status
Modified
Published
2019-03-26
Last modified
2026-06-17
Assigner
[email protected]
CVSS 3.0 ([email protected])
9.8 CRITICAL — CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
CVSS 2.0 ([email protected])
7.5 HIGH — AV:N/AC:L/Au:N/C:P/I:P/A:P
Weaknesses
CWE-287, CWE-592

Description

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

Affected configurations

• baxter sigma spectrum infusion system firmware 6.05

Related

Weakness · not in database (1)

Weakness · CWE (1)

  • CWE-287 Improper Authentication weakness

References