CVEvulnerability

CVE-2015-1012

7.5 HIGH Published 2019-03-25 Modified 2026-06-17 Official page ↗
Status
Modified
Published
2019-03-25
Last modified
2026-06-17
Assigner
[email protected]
CVSS 3.0 ([email protected])
7.5 HIGH — CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
CVSS 2.0 ([email protected])
5.0 MEDIUM — AV:N/AC:L/Au:N/C:P/I:N/A:N
Weaknesses
CWE-200, CWE-312

Description

Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.

Affected configurations

• pfizer lifecare pca infusion system firmware (<= 5.0)