CVEvulnerability
CVE-2015-1012
7.5 HIGH
Official page ↗
- Status
- Modified
- Published
- 2019-03-25
- Last modified
- 2026-06-17
- Assigner
- [email protected]
- CVSS 3.0 ([email protected])
- 7.5 HIGH — CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
- CVSS 2.0 ([email protected])
- 5.0 MEDIUM — AV:N/AC:L/Au:N/C:P/I:N/A:N
- Weaknesses
- CWE-200, CWE-312
Description
Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.
Affected configurations
• pfizer lifecare pca infusion system firmware (<= 5.0)
Related
References
- https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01 Mitigation, Third Party Advisory, US Government Resource
- https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01 Mitigation, Third Party Advisory, US Government Resource