CVEvulnerability
CVE-2015-3952
7.5 HIGH
Official page ↗
- Status
- Modified
- Published
- 2019-03-25
- Last modified
- 2026-06-17
- Assigner
- [email protected]
- CVSS 3.0 ([email protected])
- 7.5 HIGH — CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
- CVSS 2.0 ([email protected])
- 5.0 MEDIUM — AV:N/AC:L/Au:N/C:P/I:N/A:N
- Weaknesses
- CWE-200, CWE-312
Description
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Affected configurations
• pifzer plum a\+ infusion system firmware (<= 13.4)
• pifzer plum a\+3 infusion system firmware (<= 13.6)
• pifzer symbiq infusion system firmware (<= 3.13)
Related
References
- https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01 Mitigation, Third Party Advisory, US Government Resource
- https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01 Mitigation, Third Party Advisory, US Government Resource