CVEvulnerability

CVE-2015-3953

9.8 CRITICAL Published 2019-03-25 Modified 2026-06-17 Official page ↗
Status
Modified
Published
2019-03-25
Last modified
2026-06-17
Assigner
[email protected]
CVSS 3.0 ([email protected])
9.8 CRITICAL — CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
CVSS 2.0 ([email protected])
10.0 HIGH — AV:N/AC:L/Au:N/C:C/I:C/A:C
Weaknesses
CWE-259, CWE-798

Description

Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Affected configurations

• pifzer plum a\+ infusion system firmware (<= 13.4) • pifzer plum a\+3 infusion system firmware (<= 13.6) • pifzer symbiq infusion system firmware (<= 3.13)

Related

Weakness · CWE (2)

  • CWE-259 Use of Hard-coded Password weakness
  • CWE-798 Use of Hard-coded Credentials weakness

References