CVEvulnerability
CVE-2016-5085
7.5 HIGH
Official page ↗
- Status
- Modified
- Published
- 2016-10-05
- Last modified
- 2026-06-17
- Assigner
- [email protected]
- CVSS 3.0 ([email protected])
- 7.5 HIGH — CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N
- CVSS 2.0 ([email protected])
- 7.8 HIGH — AV:N/AC:L/Au:N/C:N/I:C/A:N
- Weaknesses
- CWE-330
Description
Johnson & Johnson Animas OneTouch Ping devices do not properly generate random numbers, which makes it easier for remote attackers to spoof meters by sniffing the network and then engaging in an authentication handshake.
Affected configurations
• animas onetouch ping firmware
Related
Weakness · CWE (1)
- CWE-330 Use of Insufficiently Random Values weakness
References
- http://www.kb.cert.org/vuls/id/884840 Third Party Advisory, US Government Resource
- http://www.kb.cert.org/vuls/id/BLUU-A9SQRS Third Party Advisory, US Government Resource
- http://www.securityfocus.com/bid/93351 www.securityfocus.com
- https://community.rapid7.com/community/infosec/blog/2016/10/04/r7-2016-07-multiple-vulnerabilities-in-animas-onetouch-ping-insulin-pump Mitigation, Technical Description, Third Party Advisory
- https://ics-cert.us-cert.gov/advisories/ICSMA-16-279-01 ics-cert.us-cert.gov
- http://www.kb.cert.org/vuls/id/884840 Third Party Advisory, US Government Resource
- http://www.kb.cert.org/vuls/id/BLUU-A9SQRS Third Party Advisory, US Government Resource
- http://www.securityfocus.com/bid/93351 www.securityfocus.com
- https://community.rapid7.com/community/infosec/blog/2016/10/04/r7-2016-07-multiple-vulnerabilities-in-animas-onetouch-ping-insulin-pump Mitigation, Technical Description, Third Party Advisory
- https://ics-cert.us-cert.gov/advisories/ICSMA-16-279-01 ics-cert.us-cert.gov