CVEvulnerability
CVE-2017-14018
4.8 MEDIUM
Official page ↗
- Status
- Modified
- Published
- 2017-12-05
- Last modified
- 2026-06-17
- Assigner
- [email protected]
- CVSS 3.0 ([email protected])
- 4.8 MEDIUM — CVSS:3.0/AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:L
- CVSS 2.0 ([email protected])
- 3.3 LOW — AV:L/AC:M/Au:N/C:N/I:P/A:P
- Weaknesses
- CWE-287
Description
An improper authentication issue was discovered in Johnson & Johnson Ethicon Endo-Surgery Generator Gen11, all versions released before November 29, 2017. The security authentication mechanism used between the Ethicon Endo-Surgery Generator Gen11 and single-patient use products can be bypassed, allowing for unauthorized devices to be connected to the generator, which could result in a loss of integrity or availability.
Affected configurations
• ethicon endo-surgery generator gen11 firmware
Related
Weakness · CWE (1)
- CWE-287 Improper Authentication weakness
References
- http://www.securityfocus.com/bid/101978 Third Party Advisory, VDB Entry
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-332-01 Third Party Advisory, US Government Resource
- http://www.securityfocus.com/bid/101978 Third Party Advisory, VDB Entry
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-332-01 Third Party Advisory, US Government Resource