CVEvulnerability
CVE-2017-5149
8.9 HIGH
Official page ↗
- Status
- Modified
- Published
- 2017-02-13
- Last modified
- 2026-06-17
- Assigner
- [email protected]
- CVSS 3.1 ([email protected])
- 8.9 HIGH — CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:L/I:H/A:H
- CVSS 2.0 ([email protected])
- 6.8 MEDIUM — AV:N/AC:M/Au:N/C:P/I:P/A:P
- Weaknesses
- CWE-476
Description
An issue was discovered in St. Jude Medical Merlin@home, versions prior to Version 8.2.2 (RF models: EX1150; Inductive models: EX1100; and Inductive models: EX1100 with MerlinOnDemand capability). The identities of the endpoints for the communication channel between the transmitter and St. Jude Medical's web site, Merlin.net, are not verified. This may allow a man-in-the-middle attacker to access or influence communications between the identified endpoints.
Affected configurations
• abbott merlin\@home firmware (<= 8.0)
Related
Weakness · CWE (1)
- CWE-476 NULL Pointer Dereference weakness
References
- http://www.securityfocus.com/bid/95331 Third Party Advisory, VDB Entry
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-009-01A Mitigation, Third Party Advisory, US Government Resource
- http://www.securityfocus.com/bid/95331 Third Party Advisory, VDB Entry
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-009-01A Mitigation, Third Party Advisory, US Government Resource