CVEvulnerability

CVE-2018-10596

7.1 HIGH Published 2018-07-03 Modified 2026-06-17 Official page ↗
Status
Modified
Published
2018-07-03
Last modified
2026-06-17
Assigner
[email protected]
CVSS 3.1 ([email protected])
7.1 HIGH — CVSS:3.1/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H
CVSS 3.0 ([email protected])
8.0 HIGH — CVSS:3.0/AV:A/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H
CVSS 2.0 ([email protected])
5.2 MEDIUM — AV:A/AC:L/Au:S/C:P/I:P/A:P
Weaknesses
CWE-200, CWE-923

Description

Medtronic 2090 CareLink Programmer uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.

Affected configurations

• medtronic 2090 carelink programmer firmware

Related

Weakness · CWE (2)

  • CWE-200 Exposure of Sensitive Information to an Unauthorized Actor weakness
  • CWE-923 Improper Restriction of Communication Channel to Intended Endpoints weakness

References