CVEvulnerability
CVE-2018-14781
5.3 MEDIUM
Official page ↗
- Status
- Modified
- Published
- 2018-08-13
- Last modified
- 2026-06-17
- Assigner
- [email protected]
- CVSS 3.1 ([email protected])
- 5.3 MEDIUM — CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N
- CVSS 3.0 ([email protected])
- 5.3 MEDIUM — CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N
- CVSS 2.0 ([email protected])
- 2.9 LOW — AV:A/AC:M/Au:N/C:N/I:P/A:N
- Weaknesses
- CWE-287, CWE-294
Description
Medtronic MiniMed MMT
devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
Affected configurations
• medtronicdiabetes 508 minimed insulin pump firmware
• medtronicdiabetes 522 paradigm real-time firmware
• medtronicdiabetes 722 paradigm real-time firmware
• medtronicdiabetes 523 paradigm revel firmware
• medtronicdiabetes 723 paradigm revel firmware
• medtronicdiabetes 523k paradigm revel firmware
• medtronicdiabetes 723k paradigm revel firmware
• medtronicdiabetes 551 minimed 530g firmware
• medtronicdiabetes 751 minimed 530g firmware
Related
References
- http://www.securityfocus.com/bid/105044 Third Party Advisory, VDB Entry
- https://global.medtronic.com/xg-en/product-security/security-bulletins/minimed.html global.medtronic.com
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02 Third Party Advisory, US Government Resource
- http://www.securityfocus.com/bid/105044 Third Party Advisory, VDB Entry