CVE-2018-19010

6.5 MEDIUM Published 2019-01-28 Modified 2026-06-17 Official page ↗
Status
Modified
Published
2019-01-28
Last modified
2026-06-17
Assigner
[email protected]
CVSS 3.0 ([email protected])
6.5 MEDIUM — CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
CVSS 2.0 ([email protected])
3.3 LOW — AV:A/AC:L/Au:N/C:N/I:N/A:P
Weaknesses
CWE-20

Description

Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity.

Affected configurations

• draeger kappa firmware • draeger infinity explorer c700 firmware • draeger delta xl firmware • draeger infinity delta firmware

Related

Weakness · CWE (1)

  • CWE-20 Improper Input Validation weakness