CVEvulnerability

CVE-2018-5438

6.3 MEDIUM Published 2018-03-20 Modified 2026-06-17 Official page ↗
Status
Modified
Published
2018-03-20
Last modified
2026-06-17
Assigner
[email protected]
CVSS 3.0 ([email protected])
6.3 MEDIUM — CVSS:3.0/AV:L/AC:H/PR:L/UI:N/S:U/C:H/I:H/A:N
CVSS 2.0 ([email protected])
3.3 LOW — AV:L/AC:M/Au:N/C:P/I:P/A:N
Weaknesses
CWE-613

Description

Philips ISCV application prior to version 2.3.0 has an insufficient session expiration vulnerability where an attacker could reuse the session of a previously logged in user. This vulnerability exists when using ISCV together with an Electronic Medical Record (EMR) system, where ISCV is in KIOSK mode for multiple users and using Windows authentication. This may allow an attacker to gain unauthorized access to patient health information and potentially modify this information.

Affected configurations

• philips intellispace cardiovascular (<= 2.3.0)

Related

Weakness · not in database (1)

References