CVEvulnerability
CVE-2019-13539
7.8 HIGH
Official page ↗
- Status
- Modified
- Published
- 2019-11-08
- Last modified
- 2026-06-17
- Assigner
- [email protected]
- CVSS 3.1 ([email protected])
- 7.0 HIGH — CVSS:3.1/AV:L/AC:H/PR:L/UI:N/S:U/C:H/I:H/A:H
- CVSS 3.1 ([email protected])
- 7.8 HIGH — CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H
- CVSS 2.0 ([email protected])
- 7.2 HIGH — AV:L/AC:L/Au:N/C:C/I:C/A:C
- Weaknesses
- CWE-326, CWE-328
Description
Medtronic Valleylab Exchange Client version 3.4 and below, Valleylab FT10 Energy Platform (VLFT10GEN) software version 4.0.0 and below, and Valleylab FX8 Energy Platform (VLFX8GEN) software version 1.1.0 and below use the descrypt algorithm for OS password hashing. While interactive, network-based logons are disabled, and attackers can use the other vulnerabilities within this report to obtain local shell access and access these hashes.
Affected configurations
• medtronic valleylab exchange client (<= 3.4)
• medtronic valleylab ft10 energy platform firmware (<= 4.0.0)
• medtronic valleylab fx8 energy platform firmware (<= 1.1.0)
Related
References
- https://global.medtronic.com/xg-en/product-security/security-bulletins/valleylab-generator-rfid-vulnerabilities.html global.medtronic.com
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-311-02 www.cisa.gov
- https://www.us-cert.gov/ics/advisories/icsma-19-311-02 Third Party Advisory, US Government Resource