CVEvulnerability
CVE-2019-13543
7.5 HIGH
Official page ↗
- Status
- Modified
- Published
- 2019-11-08
- Last modified
- 2026-06-17
- Assigner
- [email protected]
- CVSS 3.1 ([email protected])
- 5.8 MEDIUM — CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:L/I:N/A:N
- CVSS 3.1 ([email protected])
- 7.5 HIGH — CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
- CVSS 2.0 ([email protected])
- 5.0 MEDIUM — AV:N/AC:L/Au:N/C:P/I:N/A:N
- Weaknesses
- CWE-798
Description
Medtronic Valleylab Exchange Client version 3.4 and below, Valleylab FT10 Energy Platform (VLFT10GEN) software version 4.0.0 and below, and Valleylab FX8 Energy Platform (VLFX8GEN) software version 1.1.0 and below use multiple sets of hard-coded credentials. If discovered, they can be used to read files on the device.
Affected configurations
• medtronic valleylab exchange client (<= 3.4)
• medtronic valleylab ft10 energy platform firmware (<= 4.0.0)
• medtronic valleylab fx8 energy platform firmware (<= 1.1.0)
Related
Weakness · CWE (1)
- CWE-798 Use of Hard-coded Credentials weakness
References
- https://global.medtronic.com/xg-en/product-security/security-bulletins/valleylab-generator-rfid-vulnerabilities.html global.medtronic.com
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-311-02 www.cisa.gov
- https://www.us-cert.gov/ics/advisories/icsma-19-311-02 Third Party Advisory, US Government Resource