CVEvulnerability
CVE-2019-18248
4.3 MEDIUM
Official page ↗
- Status
- Modified
- Published
- 2020-06-29
- Last modified
- 2026-06-17
- Assigner
- [email protected]
- CVSS 3.1 ([email protected])
- 4.3 MEDIUM — CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:N
- CVSS 2.0 ([email protected])
- 3.3 LOW — AV:A/AC:L/Au:N/C:P/I:N/A:N
- Weaknesses
- CWE-319
Description
BIOTRONIK CardioMessenger II, The affected products transmit credentials in clear-text prior to switching to an encrypted communication channel. An attacker can disclose the product’s client credentials for connecting to the BIOTRONIK Remote Communication infrastructure.
Affected configurations
• biotronik cardiomessenger ii-s gsm firmware 2.20
• biotronik cardiomessenger ii-s t-line firmware 2.20
Related
Weakness · CWE (1)
- CWE-319 Cleartext Transmission of Sensitive Information weakness
References
- https://www.us-cert.gov/ics/advisories/icsma-20-170-05 Third Party Advisory, US Government Resource
- https://www.us-cert.gov/ics/advisories/icsma-20-170-05 Third Party Advisory, US Government Resource