CVEvulnerability
CVE-2019-25718
8.6 HIGH
Official page ↗
- Status
- Analyzed
- Published
- 2026-06-01
- Last modified
- 2026-07-22
- Assigner
- [email protected]
- CVSS 4.0 ([email protected])
- 8.6 HIGH — CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
- CVSS 3.1 ([email protected])
- 8.4 HIGH — CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- CVSS 3.1 ([email protected])
- 7.8 HIGH — CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H
- Weaknesses
- CWE-451
Description
Dräger Infinity Explorer C700 contains a privilege escalation vulnerability that allows attackers to break out of kiosk mode and access the underlying operating system through a specific dialog interaction. Attackers can exploit this kiosk escape to take control of the operating system and cause the device to display incorrect or no information from the connected Delta Family patient monitor.
Affected configurations
• draeger infinity explorer c700 firmware
Related
Weakness · CWE (1)
- CWE-451 User Interface (UI) Misrepresentation of Critical Information weakness