CVEvulnerability
CVE-2019-25719
8.8 HIGH
Official page ↗
- Status
- Awaiting Analysis
- Published
- 2026-06-02
- Last modified
- 2026-07-22
- Assigner
- [email protected]
- CVSS 4.0 ([email protected])
- 8.8 HIGH — CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
- CVSS 3.1 ([email protected])
- 8.6 HIGH — CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
- Weaknesses
- CWE-924
Description
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
Related
Weakness · CWE (1)
- CWE-924 Improper Enforcement of Message Integrity During Transmission in a Communication Channel weakness