CVE-2019-25724

7.1 HIGH Published 2026-06-02 Modified 2026-07-22 Official page ↗
Status
Awaiting Analysis
Published
2026-06-02
Last modified
2026-07-22
Assigner
[email protected]
CVSS 4.0 ([email protected])
7.1 HIGH — CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
CVSS 3.1 ([email protected])
6.5 MEDIUM — CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
Weaknesses
CWE-400

Description

Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.

Related

Weakness · CWE (1)

  • CWE-400 Uncontrolled Resource Consumption weakness