CVEvulnerability
CVE-2020-10627
8.1 HIGH
Official page ↗
- Status
- Modified
- Published
- 2021-12-01
- Last modified
- 2026-06-17
- Assigner
- [email protected]
- CVSS 3.1 ([email protected])
- 7.3 HIGH — CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:L
- CVSS 3.1 ([email protected])
- 8.1 HIGH — CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N
- CVSS 2.0 ([email protected])
- 4.8 MEDIUM — AV:A/AC:L/Au:N/C:P/I:P/A:N
- Weaknesses
- CWE-284
Description
Insulet Omnipod Insulin Management System insulin pump product ID 19191 and 40160 is designed to communicate using a wireless RF with an Insulet manufactured Personal Diabetes Manager device. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with access to one of the affected insulin pump models may be able to modify and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.
Affected configurations
• insulet omnipod insulin management system firmware
Related
Weakness · CWE (1)
- CWE-284 Improper Access Control weakness
References
- https://us-cert.cisa.gov/ics/advisories/icsma-20-079-01 Third Party Advisory, US Government Resource
- https://www.myomnipod.com/security-bulletins Vendor Advisory
- https://us-cert.cisa.gov/ics/advisories/icsma-20-079-01 Third Party Advisory, US Government Resource
- https://www.myomnipod.com/security-bulletins Vendor Advisory