CVEvulnerability
CVE-2020-12037
7.5 HIGH
Official page ↗
- Status
- Modified
- Published
- 2020-06-29
- Last modified
- 2026-06-17
- Assigner
- [email protected]
- CVSS 3.1 ([email protected])
- 7.5 HIGH — CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
- CVSS 2.0 ([email protected])
- 5.0 MEDIUM — AV:N/AC:L/Au:N/C:P/I:N/A:N
- Weaknesses
- CWE-259, CWE-319
Description
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
Affected configurations
• baxter prismaflex firmware
• baxter prismax firmware (< 3.0)
Related
References
- https://www.us-cert.gov/ics/advisories/icsma-20-170-01 Not Applicable
- https://us-cert.cisa.gov/ics/advisories/icsma-20-170-02 Third Party Advisory
- https://www.us-cert.gov/ics/advisories/icsma-20-170-01 Not Applicable