CVEvulnerability
CVE-2020-12039
2.4 LOW
Official page ↗
- Status
- Modified
- Published
- 2020-06-29
- Last modified
- 2026-06-17
- Assigner
- [email protected]
- CVSS 3.1 ([email protected])
- 2.4 LOW — CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:N
- CVSS 2.0 ([email protected])
- 2.1 LOW — AV:L/AC:L/Au:N/C:P/I:N/A:N
- Weaknesses
- CWE-259, CWE-798
Description
Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered on the keypad provide access to biomedical menus including device settings, view calibration values, network configuration of Sigma Spectrum WBM if installed.
Affected configurations
• baxter sigma spectrum infusion system firmware (>= 6.0 <= 6.05)
• baxter sigma spectrum infusion system firmware 8.0
Related
References
- https://www.us-cert.gov/ics/advisories/icsma-20-170-04 Third Party Advisory, US Government Resource
- https://www.us-cert.gov/ics/advisories/icsma-20-170-04 Third Party Advisory, US Government Resource