CVE-2020-15486
6.5 MEDIUM
Official page ↗
- Status
- Modified
- Published
- 2020-08-26
- Last modified
- 2026-06-17
- Assigner
- [email protected]
- CVSS 3.1 ([email protected])
- 6.5 MEDIUM — CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
- CVSS 2.0 ([email protected])
- 3.3 LOW — AV:A/AC:L/Au:N/C:P/I:N/A:N
Description
An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.
Affected configurations
• drtrust electrocardiogram pen firmware 2.00.08