CVEvulnerability

CVE-2020-16214

5.0 MEDIUM Published 2020-09-11 Modified 2026-06-17 Official page ↗
Status
Modified
Published
2020-09-11
Last modified
2026-06-17
Assigner
[email protected]
CVSS 3.1 ([email protected])
5.0 MEDIUM — CVSS:3.1/AV:L/AC:L/PR:N/UI:R/S:C/C:L/I:L/A:N
CVSS 2.0 ([email protected])
5.8 MEDIUM — AV:N/AC:M/Au:N/C:P/I:P/A:N
Weaknesses
CWE-1236

Description

In Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, the software saves user-provided information into a comma-separated value (CSV) file, but it does not neutralize or incorrectly neutralizes special elements that could be interpreted as a command when the file is opened by spreadsheet software.

Affected configurations

• philips patient information center ix b.02 • philips patient information center ix c.02 • philips patient information center ix c.03

Related

Weakness · CWE (1)

  • CWE-1236 Improper Neutralization of Formula Elements in a CSV File weakness