Description
In Patient Information Center iX (PICiX) Versions C.02, C.03,
PerformanceBridge Focal Point Version A.01, the product receives input
that is expected to be well-formed (i.e., to comply with a certain
syntax) but it does not validate or incorrectly validates that the input
complies with the syntax, causing the certificate enrollment service to
crash. It does not impact monitoring but prevents new devices from
enrolling.
Affected configurations
• philips patient information center ix b.02
• philips patient information center ix c.02
• philips patient information center ix c.03
• philips performancebridge focal point a.01
Related
Weakness · CWE (1)
- CWE-1286 Improper Validation of Syntactic Correctness of Input weakness
References
- https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01 Third Party Advisory, US Government Resource
- https://www.philips.com/productsecurity www.philips.com
- https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01 Third Party Advisory, US Government Resource
- https://www.philips.com/productsecurity www.philips.com