CVEvulnerability

CVE-2020-16228

6.4 MEDIUM Published 2020-09-11 Modified 2026-06-17 Official page ↗
Status
Modified
Published
2020-09-11
Last modified
2026-06-17
Assigner
[email protected]
CVSS 3.1 ([email protected])
6.4 MEDIUM — CVSS:3.1/AV:A/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:L
CVSS 2.0 ([email protected])
5.2 MEDIUM — AV:A/AC:L/Au:S/C:P/I:P/A:P
Weaknesses
CWE-299

Description

In Patient Information Center iX (PICiX) Versions C.02 and C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX550, MX750, MX850, and IntelliVue X3 Versions N and prior, the software does not check or incorrectly checks the revocation status of a certificate, which may cause it to use a compromised certificate.

Affected configurations

• philips patient information center ix b.02 • philips patient information center ix c.02 • philips patient information center ix c.03 • philips performancebridge focal point a.01 • philips intellivue mp2-mp90 firmware • philips intellivue mx100 firmware • philips intellivue mx400 firmware • philips intellivue mx850 firmware • philips intellivue x2 firmware • philips intellivue x3 firmware • philips intellivue mx800 firmware • philips intellivue mx750 firmware • philips intellivue mx700 firmware • philips intellivue mx600 firmware • philips intellivue mx550 firmware

Related

Weakness · CWE (1)

  • CWE-299 Improper Check for Certificate Revocation weakness