CVEvulnerability

CVE-2020-27262

5.4 MEDIUM Published 2021-01-08 Modified 2026-06-17 Official page ↗
Status
Modified
Published
2021-01-08
Last modified
2026-06-17
Assigner
[email protected]
CVSS 3.1 ([email protected])
5.4 MEDIUM — CVSS:3.1/AV:N/AC:L/PR:L/UI:R/S:C/C:L/I:L/A:N
CVSS 2.0 ([email protected])
3.5 LOW — AV:N/AC:M/Au:S/C:N/I:P/A:N
Weaknesses
CWE-79

Description

Innokas Yhtymä Oy Vital Signs Monitor VC150 prior to Version 1.7.15 A stored cross-site scripting (XSS) vulnerability exists in the affected products that allow an attacker to inject arbitrary web script or HTML via the filename parameter to multiple update endpoints of the administrative web interface.

Affected configurations

• innokasmedical vital signs monitor vc150 firmware (< 1.7.15)

Related

Weakness · CWE (1)

  • CWE-79 Improper Neutralization of Input During Web Page Generation weakness

References