CVEvulnerability
CVE-2021-33885
9.8 CRITICAL
Official page ↗
- Status
- Modified
- Published
- 2021-08-25
- Last modified
- 2026-06-17
- Assigner
- [email protected]
- CVSS 3.1 ([email protected])
- 10.0 CRITICAL — CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:N
- CVSS 3.1 ([email protected])
- 9.8 CRITICAL — CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- CVSS 2.0 ([email protected])
- 10.0 HIGH — AV:N/AC:L/Au:N/C:C/I:C/A:C
- Weaknesses
- CWE-347
Description
An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data that will be used in place of the correct data. This results in full system command access and execution because of the lack of cryptographic signatures on critical data sets.
Affected configurations
• bbraun spacecom2 (< 012u000062)
Related
Weakness · CWE (1)
- CWE-347 Improper Verification of Cryptographic Signature weakness
References
- https://www.bbraunusa.com/en.htm Broken Link
- https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/ Exploit, Third Party Advisory
- https://www.bbraunusa.com/en.htm Broken Link
- https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/ Exploit, Third Party Advisory