CVEvulnerability
CVE-2021-37160
9.8 CRITICAL
Official page ↗
- Status
- Modified
- Published
- 2021-08-02
- Last modified
- 2026-06-17
- Assigner
- [email protected]
- CVSS 3.1 ([email protected])
- 9.8 CRITICAL — CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- CVSS 2.0 ([email protected])
- 7.5 HIGH — AV:N/AC:L/Au:N/C:P/I:P/A:P
- Weaknesses
- CWE-347
Description
A firmware validation issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. There is no firmware validation (e.g., cryptographic signature validation) during a File Upload for a firmware update.
Affected configurations
• swisslog-healthcare hmi-3 control panel firmware (< 7.2.5.7)
Related
Weakness · CWE (1)
- CWE-347 Improper Verification of Cryptographic Signature weakness
References
- https://www.armis.com/PwnedPiper Broken Link
- https://www.swisslog-healthcare.com Product
- https://www.swisslog-healthcare.com/-/media/swisslog-healthcare/documents/customer-service/armis-documents/cve-2021-37160-bulletin---no-firmware-update-validation.pdf?rev=c7f94647037c4007992e2e626d445561&hash=E89531490070A809FB74994018BA1248 Vendor Advisory
- https://www.swisslog-healthcare.com/en-us/customer-care/security-information/cve-disclosures#:~:text=CVE%20Disclosures%20%20%20%20Vulnerability%20Name%20%2C%20%20CVE-2021-37164%20%204%20more%20rows%20 www.swisslog-healthcare.com
- https://www.armis.com/PwnedPiper Broken Link
- https://www.swisslog-healthcare.com Product
- https://www.swisslog-healthcare.com/-/media/swisslog-healthcare/documents/customer-service/armis-documents/cve-2021-37160-bulletin---no-firmware-update-validation.pdf?rev=c7f94647037c4007992e2e626d445561&hash=E89531490070A809FB74994018BA1248 Vendor Advisory