CVEvulnerability
CVE-2021-37162
9.8 CRITICAL
Official page ↗
- Status
- Modified
- Published
- 2021-08-02
- Last modified
- 2026-06-17
- Assigner
- [email protected]
- CVSS 3.1 ([email protected])
- 9.8 CRITICAL — CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- CVSS 2.0 ([email protected])
- 7.5 HIGH — AV:N/AC:L/Au:N/C:P/I:P/A:P
- Weaknesses
- CWE-120
Description
A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. If an attacker sends a malformed UDP message, a buffer underflow occurs, leading to an out-of-bounds copy and possible remote code execution.
Affected configurations
• swisslog-healthcare hmi-3 control panel firmware (< 7.2.5.7)
Related
Weakness · CWE (1)
- CWE-120 Buffer Copy without Checking Size of Input weakness
References
- https://www.armis.com/PwnedPiper Third Party Advisory
- https://www.swisslog-healthcare.com Product
- https://www.swisslog-healthcare.com/-/media/swisslog-healthcare/documents/customer-service/armis-documents/cve-2021-37162-bulletin---overflow-in-sccprocessmsg.pdf?rev=55a2a1d76199435688a8479970fc54bf&hash=4FDAB2F0EB319F0B773500669D67F3AD Vendor Advisory
- https://www.swisslog-healthcare.com/en-us/customer-care/security-information/cve-disclosures#:~:text=CVE%20Disclosures%20%20%20%20Vulnerability%20Name%20%2C%20%20CVE-2021-37164%20%204%20more%20rows%20 www.swisslog-healthcare.com
- https://www.armis.com/PwnedPiper Third Party Advisory
- https://www.swisslog-healthcare.com Product
- https://www.swisslog-healthcare.com/-/media/swisslog-healthcare/documents/customer-service/armis-documents/cve-2021-37162-bulletin---overflow-in-sccprocessmsg.pdf?rev=55a2a1d76199435688a8479970fc54bf&hash=4FDAB2F0EB319F0B773500669D67F3AD Vendor Advisory
- https://www.swisslog-healthcare.com/en-us/customer-care/security-information/cve-disclosures#:~:text=CVE%20Disclosures%20%20%20%20Vulnerability%20Name%20%2C%20%20CVE-2021-37164%20%204%20more%20rows%20 www.swisslog-healthcare.com