CVEvulnerability

CVE-2021-37164

9.8 CRITICAL Published 2021-08-02 Modified 2026-06-17 Official page ↗
Status
Modified
Published
2021-08-02
Last modified
2026-06-17
Assigner
[email protected]
CVSS 3.1 ([email protected])
9.8 CRITICAL — CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
CVSS 2.0 ([email protected])
7.5 HIGH — AV:N/AC:L/Au:N/C:P/I:P/A:P
Weaknesses
CWE-787

Description

A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. In the tcpTxThread function, the received data is copied to a stack buffer. An off-by-3 condition can occur, resulting in a stack-based buffer overflow.

Affected configurations

• swisslog-healthcare hmi-3 control panel firmware (< 7.2.5.7)

Related

Weakness · CWE (1)

  • CWE-787 Out-of-bounds Write weakness