CVEvulnerability
CVE-2021-37164
9.8 CRITICAL
Official page ↗
- Status
- Modified
- Published
- 2021-08-02
- Last modified
- 2026-06-17
- Assigner
- [email protected]
- CVSS 3.1 ([email protected])
- 9.8 CRITICAL — CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- CVSS 2.0 ([email protected])
- 7.5 HIGH — AV:N/AC:L/Au:N/C:P/I:P/A:P
- Weaknesses
- CWE-787
Description
A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. In the tcpTxThread function, the received data is copied to a stack buffer. An off-by-3 condition can occur, resulting in a stack-based buffer overflow.
Affected configurations
• swisslog-healthcare hmi-3 control panel firmware (< 7.2.5.7)
Related
Weakness · CWE (1)
- CWE-787 Out-of-bounds Write weakness
References
- https://www.armis.com/PwnedPiper Broken Link
- https://www.swisslog-healthcare.com Product
- https://www.swisslog-healthcare.com/-/media/swisslog-healthcare/documents/customer-service/armis-documents/cve-2021-37164-bulletin---off-by-three-stack-overflow-in-tcptxthread.pdf?rev=daf615075c71484c8059c906872a51e6&hash=1FCC1A5D921E231D71E6B95A9AA8B741 Vendor Advisory
- https://www.swisslog-healthcare.com/en-us/customer-care/security-information/cve-disclosures#:~:text=CVE%20Disclosures%20%20%20%20Vulnerability%20Name%20%2C%20%20CVE-2021-37164%20%204%20more%20rows%20 www.swisslog-healthcare.com
- https://www.armis.com/PwnedPiper Broken Link
- https://www.swisslog-healthcare.com Product
- https://www.swisslog-healthcare.com/-/media/swisslog-healthcare/documents/customer-service/armis-documents/cve-2021-37164-bulletin---off-by-three-stack-overflow-in-tcptxthread.pdf?rev=daf615075c71484c8059c906872a51e6&hash=1FCC1A5D921E231D71E6B95A9AA8B741 Vendor Advisory
- https://www.swisslog-healthcare.com/en-us/customer-care/security-information/cve-disclosures#:~:text=CVE%20Disclosures%20%20%20%20Vulnerability%20Name%20%2C%20%20CVE-2021-37164%20%204%20more%20rows%20 www.swisslog-healthcare.com