CVEvulnerability
CVE-2021-43935
9.8 CRITICAL
Official page ↗
- Status
- Modified
- Published
- 2021-12-15
- Last modified
- 2026-06-17
- Assigner
- [email protected]
- CVSS 3.1 ([email protected])
- 8.1 HIGH — CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H
- CVSS 3.1 ([email protected])
- 9.8 CRITICAL — CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- CVSS 2.0 ([email protected])
- 6.8 MEDIUM — AV:N/AC:M/Au:N/C:P/I:P/A:P
- Weaknesses
- CWE-287, CWE-288
Description
The impacted products, when configured to use SSO, are affected by an improper authentication vulnerability. This vulnerability allows the application to accept manual entry of any active directory (AD) account provisioned in the application without supplying a password, resulting in access to the application as the supplied AD account, with all associated privileges.
Affected configurations
• baxter welch allyn connex cardio (>= 1.0.0 <= 1.1.1)
• baxter welch allyn diagnostic cardiology suite 2.1.0
• baxter welch allyn rscribe resting ecg system (>= 5.01 <= 7.0.0)
• baxter welch allyn vision express holter analysis system (>= 6.1.0 <= 6.4.0)
• baxter welch allyn hscribe holter analysis system firmware (>= 5.01 <= 6.4.0)
• baxter welch allyn q-stress cardiac stress testing system firmware (>= 6.0.0 <= 6.3.1)
• baxter welch allyn xscribe cardiac stress testing system firmware (>= 5.01 <= 6.3.1)
Related
References
- https://www.cisa.gov/uscert/ics/advisories/icsma-21-343-01 Mitigation, Third Party Advisory, US Government Resource
- https://www.cisa.gov/uscert/ics/advisories/icsma-21-343-01 Mitigation, Third Party Advisory, US Government Resource