CVEvulnerability

CVE-2021-4479

6.3 MEDIUM Published 2026-06-02 Modified 2026-07-22 Official page ↗
Status
Awaiting Analysis
Published
2026-06-02
Last modified
2026-07-22
Assigner
[email protected]
CVSS 4.0 ([email protected])
6.3 MEDIUM — CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
CVSS 3.1 ([email protected])
4.0 MEDIUM — CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L
Weaknesses
CWE-1286

Description

Dräger Atlan A350 versions 1.00 up to and including 1.01 contains an improper input handling vulnerability that allows attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can transmit malformed data to overload the internal processor, gradually disrupting device operation over several hours and causing loss of data transmission, delayed display of real-time curves, and deviation between displayed airway pressure values and screen curves.

Related

Weakness · CWE (1)

  • CWE-1286 Improper Validation of Syntactic Correctness of Input weakness

References