CVEvulnerability
CVE-2022-38069
6.1 MEDIUM
Official page ↗
- Status
- Modified
- Published
- 2022-09-13
- Last modified
- 2026-06-17
- Assigner
- [email protected]
- CVSS 3.1 ([email protected])
- 4.3 MEDIUM — CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L
- CVSS 3.1 ([email protected])
- 6.1 MEDIUM — CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N
- Weaknesses
- CWE-798
Description
Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters
Affected configurations
• contechealth cms8000 firmware
Related
Weakness · CWE (1)
- CWE-798 Use of Hard-coded Credentials weakness
References
- https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01 Mitigation, Third Party Advisory, US Government Resource
- https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01 Mitigation, Third Party Advisory, US Government Resource