CVEvulnerability

CVE-2022-41627

7.6 HIGH Published 2022-10-27 Modified 2026-06-17 Official page ↗
Status
Modified
Published
2022-10-27
Last modified
2026-06-17
Assigner
[email protected]
CVSS 3.1 ([email protected])
4.8 MEDIUM — CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:L/A:H
CVSS 3.1 ([email protected])
7.6 HIGH — CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
Weaknesses
CWE-311, CWE-319

Description

The physical IoT device of the AliveCor's KardiaMobile, a smartphone-based personal electrocardiogram (EKG) has no encryption for its data-over-sound protocols. Exploiting this vulnerability could allow an attacker to read patient EKG results or create a denial-of-service condition by emitting sounds at similar frequencies as the device, disrupting the smartphone microphone’s ability to accurately read the data. To carry out this attack, the attacker must be close (less than 5 feet) to pick up and emit sound waves.

Affected configurations

• alivecor kardiamobile firmware • alivecor kardiamobile 6l firmware • alivecor kardiamobile card firmware

Related

Weakness · CWE (2)

  • CWE-311 Missing Encryption of Sensitive Data weakness
  • CWE-319 Cleartext Transmission of Sensitive Information weakness

References