CVEvulnerability

CVE-2022-4992

8.8 HIGH Published 2026-06-02 Modified 2026-07-22 Official page ↗
Status
Awaiting Analysis
Published
2026-06-02
Last modified
2026-07-22
Assigner
[email protected]
CVSS 4.0 ([email protected])
8.8 HIGH — CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
CVSS 3.1 ([email protected])
8.6 HIGH — CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
Weaknesses
CWE-345

Description

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.

Related

Weakness · CWE (1)

  • CWE-345 Insufficient Verification of Data Authenticity weakness

References