CVEvulnerability

CVE-2023-0888

7.2 HIGH Published 2023-03-13 Modified 2026-06-17 Official page ↗
Status
Modified
Published
2023-03-13
Last modified
2026-06-17
Assigner
[email protected]
CVSS 3.1 ([email protected])
4.9 MEDIUM — CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:N/I:H/A:N
CVSS 3.1 ([email protected])
7.2 HIGH — CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H
Weaknesses
CWE-94, CWE-95

Description

An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks

Affected configurations

• bbraun battery-pack sp with wifi firmware (<= 053l000092) • bbraun battery-pack sp with wifi firmware (<= 054u000092)

Related

Weakness · CWE (2)

  • CWE-94 Improper Control of Generation of Code weakness
  • CWE-95 Improper Neutralization of Directives in Dynamically Evaluated Code weakness

References