CVEvulnerability

CVE-2023-25931

6.8 MEDIUM Published 2023-03-01 Modified 2026-06-17 Official page ↗
Status
Modified
Published
2023-03-01
Last modified
2026-06-17
Assigner
[email protected]
CVSS 3.1 ([email protected])
6.4 MEDIUM — CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:H
CVSS 3.1 ([email protected])
6.8 MEDIUM — CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Weaknesses
CWE-287, CWE-620

Description

Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.

Affected configurations

• medtronic interstim x clinician a51300 • medtronic micro clinician a51200

Related

Weakness · CWE (2)

  • CWE-287 Improper Authentication weakness
  • CWE-620 Unverified Password Change weakness

References