CVEvulnerability
CVE-2023-26979
3.1 LOW
Official page ↗
- Status
- Modified
- Published
- 2023-08-03
- Last modified
- 2026-07-09
- Assigner
- [email protected]
- CVSS 3.1 ([email protected])
- 3.1 LOW — CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:L/A:N
- Weaknesses
- CWE-924
Description
Bluetens Electrostimulation Device BluetensQ device app version 4.3.15 is vulnerable to Man-in-the-middle attacks in the BLE channel. It allows attackers to decrease or increase the intensity of the stimulator by hijacking the BLE communication.
Affected configurations
• bluetens bluetensq 4.3.15
Related
Weakness · CWE (1)
- CWE-924 Improper Enforcement of Message Integrity During Transmission in a Communication Channel weakness
References
- https://www.secura.com/blog/serious-safety-impact-found-in-bluetooth-low-energy-based-medical-devices Exploit, Technical Description, Third Party Advisory
- https://www.secura.com/blog/serious-safety-impact-found-in-bluetooth-low-energy-based-medical-devices Exploit, Technical Description, Third Party Advisory