CVEvulnerability
CVE-2023-30561
6.1 MEDIUM
Official page ↗
- Status
- Modified
- Published
- 2023-07-13
- Last modified
- 2026-06-17
- Assigner
- [email protected]
- CVSS 3.1 ([email protected])
- 6.1 MEDIUM — CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:H
- CVSS 3.1 ([email protected])
- 6.1 MEDIUM — CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N
- Weaknesses
- CWE-311
Description
The data flowing between the PCU and its modules is insecure. A threat actor with physical access could potentially read or modify data by attaching a specially crafted device while an infusion is running.
Affected configurations
• bd alaris 8015 pcu firmware (<= 12.1.3)
Related
Weakness · CWE (1)
- CWE-311 Missing Encryption of Sensitive Data weakness
References
- https://www.bd.com/en-us/about-bd/cybersecurity/bulletin/bd-alaris-system-with-guardrails-suite-mx Mitigation, Vendor Advisory
- https://www.bd.com/en-us/about-bd/cybersecurity/bulletin/bd-alaris-system-with-guardrails-suite-mx Mitigation, Vendor Advisory