CVEvulnerability
CVE-2023-31222
8.8 HIGH
Official page ↗
- Status
- Modified
- Published
- 2023-06-29
- Last modified
- 2026-06-17
- Assigner
- [email protected]
- CVSS 3.1 ([email protected])
- 9.8 CRITICAL — CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- CVSS 3.1 ([email protected])
- 8.8 HIGH — CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H
- Weaknesses
- CWE-502
Description
Deserialization of untrusted data in Microsoft Messaging Queuing Service in Medtronic's Paceart Optima versions 1.11 and earlier on Windows allows an unauthorized user to impact a healthcare delivery organization’s Paceart Optima system cardiac device causing data to be deleted, stolen, or modified, or the Paceart Optima system being used for further network penetration via network connectivity.
Affected configurations
• medtronic paceart optima (< 1.12)
Related
Weakness · CWE (1)
- CWE-502 Deserialization of Untrusted Data weakness
References
- https://global.medtronic.com/xg-en/product-security/security-bulletins/paceart-optima-system.html Mitigation, Vendor Advisory
- https://global.medtronic.com/xg-en/product-security/security-bulletins/paceart-optima-system.html Mitigation, Vendor Advisory