CVEvulnerability

CVE-2023-47800

9.8 CRITICAL Published 2023-11-10 Modified 2026-06-17 Official page ↗
Status
Modified
Published
2023-11-10
Last modified
2026-06-17
Assigner
[email protected]
CVSS 3.1 ([email protected])
9.8 CRITICAL — CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Weaknesses
CWE-798

Description

Natus NeuroWorks and SleepWorks before 8.4 GMA3 utilize a default password of xltek for the Microsoft SQL Server service sa account, allowing a threat actor to perform remote code execution, data exfiltration, or other nefarious actions such as tampering with data or destroying/disrupting MSSQL services.

Affected configurations

• natus neuroworks eeg (< 8.4) • natus neuroworks eeg 8.4 • natus sleepworks (< 8.4) • natus sleepworks 8.4

Related

Weakness · CWE (1)

  • CWE-798 Use of Hard-coded Credentials weakness

References