CVEvulnerability
CVE-2025-0683
8.2 HIGH
Official page ↗
- Status
- Deferred
- Published
- 2025-01-30
- Last modified
- 2026-06-17
- Assigner
- [email protected]
- CVSS 4.0 ([email protected])
- 8.2 HIGH — CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
- CVSS 3.1 ([email protected])
- 5.9 MEDIUM — CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N
- Weaknesses
- CWE-359
Description
In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text
patient data to a hard-coded public IP address when a patient is hooked
up to the monitor. This could lead to a leakage of confidential patient
data to any device with that IP address or an attacker in a
machine-in-the-middle scenario.
Related
Weakness · CWE (1)
- CWE-359 Exposure of Private Personal Information to an Unauthorized Actor weakness
References
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01 www.cisa.gov
- https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication www.fda.gov
- https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/ www.bleepingcomputer.com
- https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor www.cisa.gov
- https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication www.fda.gov