CVEvulnerability
CVE-2026-17583
8.3 HIGH
Official page ↗
- Status
- Analyzed
- Published
- 2026-08-05
- Last modified
- 2026-08-26
- Assigner
- [email protected]
- CVSS 4.0 ([email protected])
- 8.3 HIGH — CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:N/VI:H/VA:H/SC:N/SI:H/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
- CVSS 3.1 ([email protected])
- 8.4 HIGH — CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- Weaknesses
- CWE-353
Description
The affected
Thermo Fisher Applied Biosystems Genetic Analyzers are vulnerable because .fsa/.hid output files can be edited. An attacker could tamper with these files, altering DNA data and resulting in inaccurate DNA test outcomes.
Affected configurations
• thermofisher abi prism 310 data collection software (<= 3.1)
• thermofisher abi prism 3100\/3100-avant data collection software (<= 2.0)
• thermofisher applied biosystems 3130 series data collection software (<= 4.1)
• thermofisher applied biosystems 3500\/3500xl series data collection software (< 4.0.3)
• thermofisher applied biosystems 3730\/3730xl series data collection software (< 5.0.3)
• thermofisher applied biosystems seqstudio flex series instrument software (< 1.2.1)
• thermofisher applied biosystems seqstudio genetic analyzer data collection software (< 1.2.6)
• thermofisher genemapper id-x (< 1.7.4)
Related
Weakness · CWE (1)
- CWE-353 Missing Support for Integrity Check weakness
References
- https://documents.thermofisher.com/TFS-Assets/CORP/Product-Guides/fsa_hid_bulletin.pdf Patch, Vendor Advisory
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-216-01 Patch, Third Party Advisory, US Government Resource
- https://www.cve.org/CVERecord?id=CVE-2026-17583 Third Party Advisory