CVEvulnerability
CVE-2026-18844
7.2 HIGH
Official page ↗
- Status
- Deferred
- Published
- 2026-08-11
- Last modified
- 2026-09-08
- Assigner
- [email protected]
- CVSS 4.0 ([email protected])
- 7.2 HIGH — CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
- CVSS 3.1 ([email protected])
- 8.1 HIGH — CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:H
- Weaknesses
- CWE-912
Description
The firmware of the Pulsetto Vagus Nerve Stimulator accepts several undisclosed commands over its Bluetooth Low Energy (BLE) interface. These commands are sent without authentication or encryption, and are never issued by the companion mobile application, yet are fully processed by the device when it is powered on.
Related
Weakness · CWE (1)
- CWE-912 Hidden Functionality weakness