CVEvulnerability

CVE-2026-7251

9.3 CRITICAL Published 2026-05-26 Modified 2026-07-24 Official page ↗
Status
Awaiting Analysis
Published
2026-05-26
Last modified
2026-07-24
Assigner
[email protected]
CVSS 4.0 ([email protected])
9.3 CRITICAL — CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
CVSS 3.1 ([email protected])
9.8 CRITICAL — CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Weaknesses
CWE-259

Description

Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.

Related

Weakness · CWE (1)

  • CWE-259 Use of Hard-coded Password weakness

References