CVEvulnerability

CVE-2026-85628

7.0 HIGH Published 2026-09-16 Modified 2026-09-18 Official page ↗
Status
Deferred
Published
2026-09-16
Last modified
2026-09-18
Assigner
539080bd-5750-4cce-b30b-eed9a4ef6dcc
CVSS 4.0 (539080bd-5750-4cce-b30b-eed9a4ef6dcc)
7.0 HIGH — CVSS:4.0/AV:A/AC:L/AT:P/PR:N/UI:P/VC:H/VI:N/VA:N/SC:H/SI:L/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
Weaknesses
CWE-319

Description

Transmission of the home Wi-Fi credentials without encryption during the pairing process between the DuoxMe application and VEO and VEO-XS Wi-Fi monitors, in versions prior to 4.3.4 of the application and 01.50.001 of the monitor firmware, allows an attacker on the Wi-Fi Direct network to intercept the network password.

Related

Weakness · CWE (1)

  • CWE-319 Cleartext Transmission of Sensitive Information weakness

References