CVEvulnerability

CVE-2020-7590

6.8 MEDIUM Published 2020-10-13 Modified 2026-06-17 Official page ↗
Status
Modified
Published
2020-10-13
Last modified
2026-06-17
Assigner
[email protected]
CVSS 3.1 ([email protected])
6.8 MEDIUM — CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
CVSS 2.0 ([email protected])
4.6 MEDIUM — AV:L/AC:L/Au:N/C:P/I:P/A:P
Weaknesses
CWE-259

Description

A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797). Affected devices use a hard-coded password to protect the onboard database. This could allow an attacker to read and or modify the onboard database. Successful exploitation requires direct physical access to the device.

Affected configurations

• siemens dca vantage analyzer firmware (< 4.5.0.0)

Related

Weakness · CWE (1)

  • CWE-259 Use of Hard-coded Password weakness

References