CVE-2025-14097
7.2 HIGH
Official page ↗
- Status
- Deferred
- Published
- 2025-12-17
- Last modified
- 2026-06-17
- Assigner
- 46b595e9-1acc-41cb-9398-adaf98d37a9b
- CVSS 3.1 (46b595e9-1acc-41cb-9398-adaf98d37a9b)
- 7.2 HIGH — CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H
- Weaknesses
- CWE-287
Description
A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software. Other related CVE's are CVE-2025-14095 & CVE-2025-14096. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.
Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer. Temporary work Around: If the network is not considered secure, please remove the analyzer from the network. Permanent solution:
Customers should ensure the following:
• The network is secure, and access follows best practices.
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.
Exploit Status:
Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.
Related
Weakness · CWE (1)
- CWE-287 Improper Authentication weakness
References
- https://www.radiometer.com/myradiometer www.radiometer.com